Abstract

ObjectiveTo explore the direct measure of EEG amplitude (range EEG, rEEG) for detection of interburst intervals (IBIs) and bursts in neonates. MethodsPreviously described 177 two-channel EEG recordings 3–6h long from 26 preterm infants (median gestational age of 26weeks) at 23–38weeks post-menstrual age (PMA) without major abnormalities were used to test four definitions of IBI detection algorithms with various settings of the parameters. ResultsAs the basis for all four algorithms we developed the aggregation of rEEG signal over the channels by taking its maximum, and method of EEG trace selection at different phases of sleep–wake cycle (with different degree of discontinuity). The two less restrictive algorithms – with one and two amplitude thresholds - turned to be the most promising definitions. There were enough IBI detections for analysis, with no substantial difference in mean and maximum values of intervals. The longest IBI were measured at the location of greater discontinuity. Values of bursts and IBI indices as well as association with PMA were close to the published normative values derived manually. ConclusionsrEEG as a direct measure of EEG amplitude can be used for detection of bursts and IBI. SignificanceThe automatic measurement of IBI based on rEEG provides a basis for improvements in neonatal brain monitoring.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call