Abstract

In this work, an evaluation of commercially available polystyrene-divinylbenzene sorbents for solid-phase extraction (SPE) of eleven phenolic compounds is intended. Considering the particle size and cross-linking degree, Amberlite XAD-4 (commercial or grounded), Macronet MN-200 and Lichrolut EN were tested. The SPE protocol was performed by an automatic system, providing repeatable experimental conditions for assessment of sorbent capacity, breakthrough volume and enrichment factor (EF). A positive correlation between EF and log K ow was found for Amberlite XAD-4 while a negative correlation was observed between EF and molecular weight of analyte for Macronet MN-200 and for Lichrolut EN. This indicates a prevalence of hydrophobic interactions or molecular exclusion depending upon the polymer cross-linking degree. Despite the similar repeatability (RSD < 4.7%, n > 6) and recovery values attained (97.6–102.7%, using 50 mL of sample) for all sorbents, Lichrolut EN is the best choice for analytical application as higher EF and lower LOD values (between 18 and 207 ng) were attained for this sorbent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.