Abstract

Automatic exposure control (AEC) is an important feature in mammography. It enables consistently optimal image exposure despite variations in tissue density and thickness, and user skill level. Full field digital mammography systems cannot employ conventional AEC methods because digital receptors fully absorb the x-ray beam. In this paper we describe an AEC procedure for slot scanning mammography. With slot scanning detectors, our approach uses a fast low-resolution and low-exposure prescan to acquire an image of the breast. Tube potential depends on breast thickness, and the prescan histogram provides the necessary information to calculate the required tube current. We validate our approach with simulated prescan images and phantom measurements. We achieve accurate exposure tracking with thickness and density, and expect this method of AEC to reduce retakes and improve workflow.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.