Abstract

Background. Optimization of cardiac resynchronization therapy (CRT) settings after implant can improve response to therapy. In this Italian single-center experience, we investigated the rate of hemodynamic and clinical response in heart failure patients treated with continuously and automatically optimized CRT. Methods. Patients were selected from June 2015 to April 2017 according to the most recent CRT guidelines; all were in sinus rhythm at implant and received a CRT-defibrillator system equipped with SonR, which automatically optimizes AV and VV delays every week. SonR was activated just after implant and remained active during follow-up. The rate of hemodynamic response (R-HR) was defined as ΔLVEF>5%, super-response (R-HSR) as ΔLVEF>15%, and clinical response as a negative transition of NYHA class≥−1 at 6 months follow-up vs. baseline (preimplant). Results. Mean follow-up for the 31 patients (aged 69.9±9.4 years; 61% male; NYHA class II/III 19%/81%; ischemic etiology 65%) was 6±0.7 months. At baseline, LVEF was 29.1%±4.7% and QRS duration 146±13 ms. LBBB morphology was observed in 65%. At 6 months, R-HR was 74% (23/31), R-HSR 32% (10/31), and clinical response rate 77% (24/31). Hemodynamically, patients with ischemic etiology benefited more than those without ischemic etiology, both in terms of response (80% versus 64%) and super-response (35% versus 27%). Conclusions. Continuous automatic weekly optimization of CRT over 6 months consistently improved R-HR, R-HSR, and clinical response in NYHA class II/III heart failure patients versus baseline. Patients with ischemic etiology in particular may benefit hemodynamically from this type of CRT optimization.

Highlights

  • Cardiac resynchronization therapy (CRT) is a wellestablished therapy for patients with medically refractory heart failure, left ventricular (LV) systolic dysfunction, and a wide QRS complex [1]

  • New York Heart Association (NYHA) class II/III was found in 19%/81% of patients, respectively, mean LVEF was 29.1%, mean QRS duration was 146 ms, and LBBB was found in 65% of patients

  • cardiac resynchronization therapy (CRT) delivered with continuous weekly optimization of AV and VV delays based on sensorguided CRT optimization approach (SonR)-detected cardiac contractility was shown to be effective at improving the rates of response to therapy in NYHA class II/III heart failure patients after 6 months

Read more

Summary

Introduction

Cardiac resynchronization therapy (CRT) is a wellestablished therapy for patients with medically refractory heart failure, left ventricular (LV) systolic dysfunction, and a wide QRS complex [1]. As a matter of fact, several studies have demonstrated the acute hemodynamic benefits of AV and VV timing optimization guided by echocardiography [6]. Optimization of cardiac resynchronization therapy (CRT) settings after implant can improve response to therapy. In this Italian single-center experience, we investigated the rate of hemodynamic and clinical response in heart failure patients treated with continuously and automatically optimized CRT. The rate of hemodynamic response (R-HR) was defined as ΔLVEF > 5%, super-response (R-HSR) as Δ LVEF > 15%, and clinical response as a negative transition of NYHA class≥−1 at 6 months follow-up vs baseline (preimplant). Continuous automatic weekly optimization of CRT over 6 months consistently improved R-HR, R-HSR, and clinical response in NYHA class II/III heart failure patients versus baseline. Patients with ischemic etiology in particular may benefit hemodynamically from this type of CRT optimization

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.