Abstract

AbstractThe theory of an improved coulometric stripping potentiometry method, involving automatic determination of the chemical current for each sample, and thus suitable for calibration‐free determination of trace elements in microliter samples is described. The method has been applied to the determination of ZnII, CdII, PbII, TlI, CuII and BiIII by reductive in situ amalgamation and of AsV, SnIV and CuII by reductive in situ dissolution on a gold film followed by three consecutive constant current stripping potentiometry measurements permitting the determination of the chemical current. Quantitative reduction (≥99%) of the analyte ions was obtained after four minutes of vibrationally enhanced reductive electrolysis of the 12μL samples placed on top of an inverted glassy carbon disk electrode combined with Ag and Ag/ AgCl counter and reference electrodes. The concentration detection limits ranged from 11 nM for elements oxidized to the monovalent state to 3 nM for elements oxidized to the trivalent state. The mass detection limits ranged from 4.5 pg for ZnII to 28pg for TlI.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.