Abstract

Conventional bridge circuits are balanced by means of mechanical changes of trimming capacitances and resistances. It is a well known fact that this can be very difficult in many situations, e.g. due to electrical interference between the bridge circuit and the human body or some mechanical part of the screwdriver used. There may also be physical constraints prohibiting mechanical changes of some bridge elements. To avoid these problems, the balancing procedure should be automatized. In this article an automatic method for bridge circuit balancing, based on a digital control signal, is presented. Ordinary ac-bridge circuits are nonlinear with respect to changes of the capacitive sensor element. Using some standard operational amplifiers it is easy, however, to obtain a linear bridge circuit output. A theoretical model of the linear bridge circuit is derived. The excellent performance of the linear bridge circuit predicted by theory is verified by practical experiments. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.