Abstract

This study evaluated the accuracy and reliability of the Dinamap 8100 automated blood pressure machine against three internationally recognized criteria. Systolic and diastolic blood pressures were taken concurrently by two nurses using the automated machine and a manual sphygmomanometer. Results demonstrated agreement between automated and manual readings on one set of criteria for both systolic and diastolic pressures, and support for systolic readings only on one other criterion. Comparison of mean differences between automated and manual measures showed the automated machine consistently under-read both systolic and diastolic blood pressures. The conclusion from this study was that the Dinamap 8100 machine can be used with some degree of confidence to assess systolic blood pressures in a general population of adult hospital inpatients, but with caution when taking diastolic readings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.