A low-cost (under $12,000), fully automated, relative humidity calibration chamber capable of testing up to six probes simultaneously over a range of humidifies from 5% to approximately 95% is designed, constructed, and tested for use by the Oklahoma mesonetwork. In addition, a low-cost (under $4,000) temperature calibration chamber is developed that can test seven probes over the entire meteorological range of interest.(≈ −30° to +50°C). These procedures were designed and developed for the Vaisala HMP 35C temperature and relative humidity probe but are applicable to any similar instrument. The initial implementation results are presented. The relative humidity calibration uses an optical condensation dewpoint hygrometer as the laboratory reference. Two precision resistance temperature detectors are used in the temperature calibration. It should be noted that commercial temperature and relative humidity chambers are available that would have been sufficiently accurate. However, the systems developed here lend themselves to automation and can calibrate multiple probes simultaneously. The facilities developed herein are capable of producing probes with an uncertainty of ±3% for relative humidifies from 5% to approximately 90%, and an error of ±3.5% for humidifies from 90% to 95%. An uncertainty of ±0.2°C can be attained using the temperature calibration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.