An automated high-performance liquid chromatography (HPLC) assay for ethinyl estradiol and norgestrel or levonorgestrel in oral contraceptive tablets was developed. Tablets were prepared for on-line injection using a solid sampler and segmented continuous flow techniques. The active components were separated from tablet excipients, impurities, and degradation products on reversed-phase C8 and C18 columns by elution with water-acetonitrile-methanol (45:35:15). A UV detector connected in series with a fluorometric detector measured the UV absorbance of levonorgestrel and norgestrel at 240 nm and the fluorescence of ethinyl estradiol at 310 nm (excitation at 210 nm). The method employed computer control of the injection system and solid sampler for synchronization of the chromatographic and segmented flow streams. The method is applicable for content uniformity and stability testing at a rate of eight samples per hour.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.