Abstract

Anti-cardiolipin and β2-glycoprotein I antibodies represent important diagnostic parameters in routine hematology. In this study, five different automated, semi-automated, and manual immunoassays detecting IgG/IgM anti-cardiolipin and anti-β2 -glycoprotein I antibodies were tested. A total of 162 samples from women with a history of miscarriage were recruited from 110 different G&O outpatient centers in Germany. For both anti-cardiolipin and anti-β2 -glycoprotein I antibodies, considerable differences in the percentage of positive results were seen between all five methods, and itemization of all positive test results revealed a poor accordance. These findings were confirmed by Cohen's kappa coefficients. Our study revealed a moderate to poor accordance between five different test systems for anti-cardiolipin and anti-β2 -glycoprotein I antibodies. Such deviations may result in clinical misinterpretation of data and may lead to wrong therapeutic consequences. Therefore, further standardization of all tests for anti-phospholipid antibodies should be achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.