Abstract

Radioactive krypton-85 (85Kr) gas, a chemically inert and non-proliferation indicator, is derived from fission products. Its detection relies on the Budesamt für Strahlenschutz–Institute of Atmospheric Radioactivity (BfS-IAR) method, which necessitated impurity removal using soda lime, silica gel, and liquid nitrogen for cryogenic adsorption. This manual process requires frequent replacements, posing challenges for its automation. To address this, we developed a prototype krypton sampling system as an interim research product for the fully automated remote monitoring of covert nuclear activity. The system incorporates a hollow fiber membrane for impurity removal, a computer-controlled multi-position valve for sampling, and an electric cooler for adsorption. The impurity removal modules demonstrated high efficiencies, removing H2O and CO2 at 99.8% and 97.8% rates, respectively. Further, the custom-made sampling system can process 16 samples in a single run without analyst intervention. We conducted experiments to verify the automatic krypton sampling capability. The activity concentration of 85Kr in ambient air was measured using the BfS-IAR processing and detection system. The system exhibited a recovery rate of ~7.8% for krypton in 1000 L air, demonstrating good continuous remote monitoring capability. This study promotes the development of an automated analysis system for the detection of 85Kr in ambient air.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call