Abstract

An automated sample preparation system has been developed for the determination of cholesterol in a wide range of matrices. Isolation of cholesterol is performed with a robotic arm coupled with a series of modular stations. Samples are introduced into the system which adds the appropriate reagents, carries out the saponification, pH adjustment, solid-phase extraction and drying steps. This system was evaluated using 15 different food matrices. The average recovery for NIST standards exceeded 97%. A solution of n-hexane-2-propanol was substituted for the traditional methanol-chloroform extraction. Manual pH adjustment was replaced with a buffer. Manual and automated methods were compared and no difference was observed at the 95% confidence level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.