Abstract

ECG-based gating in cardiac MR imaging requires additional patient preparation time, is susceptible to RF and magnetic interference, and is ineffective in a significant percentage of patients. "Wireless" or "self-gating" techniques have been described using either interleaved central k-space lines or projection reconstruction to obtain MR signals synchronous with the cardiac cycle. However, the interleaved, central line method results in a doubling of the acquisition time, while radial streak artifacts are encountered with the projection reconstruction method. In this work, a new self-gating technique is presented to overcome these limitations. A retrospectively gated TrueFISP cine sequence was modified to acquire a short second echo after the readout and phase gradients are rewound. The information obtained from this second echo was used to derive a gating signal. This technique was compared to ECG-based gating in 10 healthy volunteers and shown to have no significant difference in image quality. The results indicate that this method could serve as an alternative gating strategy without the need for external physiological signal detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.