Abstract

¹²³I-FP-CIT (DaTSCAN) SPECT studies of the nigrostriatal pathway are a valuable tool in the diagnosis of movement disorders. However some scans are reported as equivocal with potential adverse consequences. We investigated whether the use of quantification of tracer uptake within the striatum can be used to reduce the number of equivocal reports. BRASS software (Hermes, Sweden) was used to quantify striatal tracer uptake in DaTSCAN studies of patients referred to our institution. Scans were quantified and numerical limits were determined to distinguish between normal and abnormal scans. Scans were then re-reported both with, and without, the use of quantification. Number of equivocal reports and accuracy of reporting between the two types of reporting were compared. Scan reporting using quantification led to a significant reduction in the number of equivocal reports with no significant change in reporting accuracy. Automated quantification of DaTSCAN studies with BRASS and the use of numerical limits can decrease the number of equivocal reports without affecting report accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.