Abstract

The interest in gallium-68 labelled positron-emission tomography probes continues to increase around the world. However, one of the barriers for routine clinical use is the cost of the automated synthesis units for relatively simple labelling procedures. Herein, we describe the adaptation of a TRACERlab FXFN synthesis module for the automated production of gallium-68 radiopharmaceuticals using a cation-exchange cartridge for postprocessing of the 68 Ge/68 Ga generator eluate. The recovery of activity from the cartridge was 95.6% to 98.9% using solutions of acidified sodium chloride (5M with pH=1-3). The radiosyntheses of [68 Ga]Ga-DOTANOC and [68 Ga]Ga-PSMA-11 were performed using acetate sodium buffer or 4-(2-hydroxyethyl)piperazine-1-ethanesulfonic acid, with a total duration of 21 and 23minutes, respectively, including generator elution and radiopharmaceutical dispensing. Activity yields were 77%±2% for [68 Ga]Ga-PSMA-11 and 68%±3% for [68 Ga]Ga-DOTANOC (n>100). The labelled peptides had a radiochemical purity exceeding 97%, and all quality control parameters were in conformity with the limits prescribed by the European Pharmacopoeia.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.