Abstract

Treatment planning for whole‐brain radiation treatment is technically a simple process, but in practice it takes valuable clinical time of repetitive and tedious tasks. This report presents a method that automatically segments the relevant target and normal tissues, and creates a treatment plan in only a few minutes after patient simulation. Segmentation of target and critical structures is performed automatically through morphological operations on the soft tissue and was validated by comparing with manual clinical segmentation using the Dice coefficient and Hausdorff distance. The treatment plan is generated by searching a database of previous cases for patients with similar anatomy. In this search, each database case is ranked in terms of similarity using a customized metric designed for sensitivity by including only geometrical changes that affect the dose distribution. The database case with the best match is automatically modified to replace relevant patient info and isocenter position while maintaining original beam and MLC settings. Fifteen patients with marginally acceptable treatment plans were used to validate the method. In each of these cases the anatomy was accurately segmented, but the beams and MLC settings led to a suboptimal treatment plan by either underdosing the brain or excessively irradiating critical normal tissues. For each case, the anatomy was automatically segmented with the proposed method, and the automated and manual segmentations were then compared. The mean Dice coefficient was 0.97, with a standard deviation of 0.008 for the brain, 0.85±0.009 for the eyes, and 0.67±0.11 for the lens. The mean Euclidian distance was 0.13±0.13 mm for the brain, 0.27±0.31 for the eye, and 2.34±7.23 for the lens. Each case was then subsequently matched against a database of 70 validated treatment plans and the best matching plan (termed autoplanned), was compared retrospectively with the clinical plans in terms of brain coverage and maximum doses to critical structures. Maximum doses were reduced by a maximum of 8.37 Gy for the left eye (mean 2.08), 11.67 for the right eye (1.90) and, respectively, 25.44 (5.59) for the left lens and 24.40 (4.85) for the right lens. Time to generate the autoplan, including the segmentation, was 3−4 min. Automated database‐ based matching is an alternative to classical treatment planning that improves quality while providing a cost‐effective solution to planning through modifying previous validated plans to match a current patient's anatomy.PACS number: 87.55.D, 87.55.tg, 87.57.nm

Highlights

  • Information available in the form of previous clinical cases used to treat similar anatomy is increasingly used in radiotherapy to judge the quality of new plans or to suggest constraints for the inverse planning process

  • Termed knowledge-based treatment planning, this information becomes largely available with the informatization of radiotherapy practice in the comprehensive database systems that are the basis for modern software in the treatment planning process

  • We introduce an efficient method to segment the relevant anatomy for whole-brain radiation therapy and search a validated clinical database for prior cases with similar anatomy that could without modification be used for treatment

Read more

Summary

Introduction

Information available in the form of previous clinical cases used to treat similar anatomy is increasingly used in radiotherapy to judge the quality of new plans or to suggest constraints for the inverse planning process. Geometry-based approaches that consider individual patient topology have been used to create maps that describe a specific direction to access a tumor without intersecting critical organs.[10,11] When incorporated into the treatment planning process,(3,4,12,13) this a priori information improves planning times as the algorithm employs favorable beam settings from the geometrical location of critical structures relative to the PTV

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.