Abstract

The assay for C-reactive protein has been fully automated as a particle-counting immunoassay. For cerebrospinal fluid, cord serum, and adults' serum its range of sensitivity extends from 1 microgram/L to 300 mg/L, with a minimal sample dilution of twofold and a maximal dilution of 50-fold being required. This range is so broad because free antibodies are added to the reaction medium. However, we have used Fab fragments rather than whole antibody to avoid too steep a standard curve and a decrease of agglutination at high concentrations of antigen. For 99 consecutive cord sera examined, the concentrations of C-reactive protein ranged from 7 micrograms/L to 1.750 mg/L. The geometric mean was 50 micrograms/L and the upper normal limit (geometric mean +/- 2 SD of the log values) was established at 525 micrograms/L.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.