Abstract

A flow injection on-line sorption preconcentration system for the electrothermal atomic absorption spectrometric determination of platinum has been developed. The method employs on-line preconcentration of platinum on a column packed with silica gel functionalised with 1,5-bis(di-2-pyridyl)methylene thiocarbohydrazide (DPTH-gel) placed in the autosampler arm. The modification of the autosampler in the tubing line and circuit allowed either the flow of the sample through the column or the operation of the autosampler in the normal mode, where microlitres of 2 M HNO 3, which acts as the elution agent, pass through the microcolumn eluting Pt(IV), which is directly deposited in the graphite tube as a drop of a precisely defined volume. An enrichment factor of 41.7 and a 3 σ detection limit of 0.8 ng ml −1 along with a sampling frequency of 29 h −1 are achieved with 60 s preconcentration time at a sample flow rate of 2.4 ml min −1. The relative standard deviation is 1.6% for 4 ng ml −1 Pt ( n=11). The method has been applied to the determination of platinum in catalyst, vegetation, soil and water samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.