Abstract
Insertion loss is an important parameter used to characterize passive fiber-optic components, such as WDM couplers and variable optical attenuators. This article describes an automated insertion loss measurement system which incorporates 980 nm, 1310 nm, 1529 nm and 1561 nm DFB lasers and two internal fiber optic standards. Insertion loss measurements collected with the system for WDM couplers and internal standards are presented. The system repeatability was validated by measuring the insertion loss for a WDM coupler six times at 980 nm, 1310 nm, 1529 nm and 1561 nm over 3 days period. The standard deviation calculated for the insertion loss measurements is less than 0.11 dB and the %CV is less than 1%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Measurement Technologies and Instrumentation Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.