Abstract

Abstract This study developed a simple wire phantom and an algorithm to automatically measure the modulation transfer function (MTF) in computed tomography (CT) and implemented it to evaluate the effect of focal spot size and reconstruction filter type. The phantom consisted of a resin cylinder filled with water, with a tin wire of diameter 0.1 mm positioned along the center of the cylinder. The automated MTF algorithm used an axial image of the phantom and comprised several steps. The center position of a region of interest (ROI) was automatically determined at the center of the wire image. The pixels were then summed along the y-direction to obtain the profile of the pixel values at a point along the x-direction. Following this, both edges of the profile were made equal to zero. The profile curve was then normalized so that the total of all the data was equal to unity. The normalized profile curve is the line spread function (LSF), and the MTF curve was obtained by taking its Fourier transform. Our system (phantom and algorithm) is able to differentiate the MTFs of CT images from different focal sizes and reconstruction filter types.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.