Abstract
This work presents a multisyringe flow injection analysis (MSFIA) system for the automatic extraction and determination of tetracyclines in wastewater samples. The sample was adjusted with Na2EDTA buffer before solid-phase extraction with an Oasis HLB column used for the analyte preconcentration. The europium (Eu3+)-based and citrate-mediated method (using Tris-HCl buffer) was selected for the fluorimetric analysis (λexc/em = 400/612 nm). For fluorescence detection, a low-cost system consisting of an USB 2000 CCD detector and a 3D-printed support that holds a LED light source was used. Under optimized conditions, the proposed method provided low limits of detection (9.4 μg L−1) and quantification (31 μg L−1), and good values for intra-day (<4 %) and inter-day precisions (<6 %). Recoveries of spiked TCs in wastewater samples ranged from 87 to 106 %. The results of this work were in good agreement with the measurements obtained by liquid chromatography coupled to a fluorescence detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.