A method for the determination of angiotensin-converting enzyme in serum (S-ACE; EC 3.4.15.1.) with use of 3-(2-furylacryloyl)-L-phenylalanyl-glycyl-glycine (FAPGG) as substrate has been adapted for the Cobas Bio microcentrifugal analyser. The method allows 24 determinations per hour in a sample volume of 28 microliters with a within run precision of less than 3% and a between run precision of less than 5%. The reaction is linear up to at least 470 U/l. The reference interval in 92 blood donors has been determined to 14-110 U/l. The method correlates well with the manual method of Hurst & Lovell-Smith (r = 0.982). We have found the method excellently suited for routine assay.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scandinavian journal of clinical and laboratory investigation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.