Abstract

Abstract In response to our need for a rapid method of analyzing numerous aluminum samples, we developed an automated procedure for measuring total monomeric aluminum in aqueous samples. Two separate designs for a continuous microflow analyzer are presented which enable detection over a combined range of 20 μg Al/l to 1000 μg Al/l. The best fit to the calibration curve for the high sensitivity manifold is not linear (r2 = 0.998) as previously reported but third order (cubic, r2 = 1.000). The choice of equations for curve fit for the low sensitivity range is less critical as first and third order equations provide comparable fit. Between-sample contamination is less that 3% and reproducibility is high; less than 1% variation for high (200 - 1000 μg Al/l) concentrations and less than 7% for lower (20 - 300 μg Al/l concentrations. Analysis of both water and soil extract samples for aluminum should be facilitated by use of this system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.