A candidate reference measurement procedure involving automated isotope dilution coupled with liquid chromatography–tandem mass spectrometry (ID-LC–MS/MS) with on-line dilution and solid phase extraction (SPE) has been developed and critically evaluated. We constructed the LC–MS/MS with on-line dilution and SPE system. An isotopically labelled internal standard, cortisol-d4, was added to serum sample. After equilibration, the methanol was added to the sample, and deproteination was performed. Then, the sample was applied to the LC–MS/MS system. The limit of detection (LOD) and limit of quantification (LOQ) were 0.2 and 1ngg−1, respectively. Excellent precision was obtained with within-day variation (RSD) of 1.9% for ID-LC–MS/MS analysis (n=6). This method, which demonstrates simple, easy, good accuracy, high precision, and is free from interferences from structural analogues, qualifies as a reference measurement procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.