Abstract

The injection of a radioactive reaction mixture into a preparative HPLC column has been automated with computer control for rapid purification of routinely prepared positron emitting radiopharmaceuticals. Using pneumatic valves, a motor-driven pump and a liquid level sensor, two intelligent injection methods for the automation were compared with regard to efficient and rapid sample loading into a 2 mL loop of the 6-way valve. One, a precise but rather slow method, was demonstrated to be suitable for purification of 18F-radiopharmaceuticals, while the other, due to its rapid operation, was more suitable for 11C-radiopharmaceuticals. A sample volume of approx 0.5 mL can be injected onto a preparative HPLC column in over 90% efficiency with the present automated system.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.