Abstract
A computer-controlled immunological binding assay intended for monitoring concentrations of interesting metabolites during bioprocesses has been developed. The main focus has been on speed, reliability, reproducibility and operational stability of the assay. The assay is based on the repeated use of a preparation of immobilized antibodies. Stability in the regeneration step is a prerequisite and was achieved by intermittent recalibration of the system. Furthermore, before immobilizing the polyclonal antibodies, they were purified on immobilized antigen and eluted using the same procedure as used in the assay cycle. Antibody preparation could be used for more than 50 cycles and in some cases for several hundred cycles. Since the calibration curve is valid for the whole life-time of the antibody column, recalibration only involved the registration of the column capacity. Experiments have been performed concerning monitoring of chromatographic separations as well as adsorption experiments from complex mixtures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.