Abstract

We developed an automated homogeneous immunoassay, based on immune lysis of dinitrophenyl (DNP)-labeled liposomes, for measuring total complement activity. Liposome lysis caused by complement activity was detected spectrophotometrically from entrapped glucose-6-phosphate dehydrogenase activity. Complement activity in human sera was quantified by comparison with a calibration curve. For ease of application to fully automated routine clinical analyzers, we adopted a two-reagent system, one reagent containing a homogeneous population of small DNP-labeled liposomes and one containing antibody/substrate. This system required calibration only once a week. Within-run and between-run CVs were 0.4-1.3% (n = 10) and 1.8-4.7% (n = 10), respectively. Serum results were linear upon dilution (with saline) over a twofold range. Bilirubin, hemoglobin, Intrafat, and serum proteins such as rheumatoid factor, M protein, IgG, and IgA did not affect the assay results. The results (y) correlated well with those from a hemolytic complement activity test (x): y = 1.05x - 1.14, r = 0.92, on 66 samples in the range < 10- > 50 kU/L. This method should therefore be of great use for the determination of complement activity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.