Abstract

An automated analyte electroextraction and preconcentration system, which was used as the front end for a liquid chromatography-electrospray mass spectrometry instrument, is described. The extraction was based on driving the anionic analytes across a polymer inclusion membrane by application of a potential of 200 V to the cell. Five milliliters (5 mL) of sample were passed through a flow-through cell at a flow rate of 0.2 mL/min containing a membrane 20 μm thick. This consisted of 75% cellulose triacetate as base polymer, 12.5% of tris(2-ethylhexyl)phosphate as plasticizer, and 12.5% of Aliquat 336 as cationic carrier. The target analytes were enriched in 20 μL of a stagnant acceptor solution prior to online LC/MS analysis. The performance of the system was demonstrated for the determination of chlorinated phenoxyacetic acid herbicides in spiked river water. Enrichment factors of ~200 were achieved with recoveries of typically 99% and precision values of typically 5%. The limit of detection (LOD) values were found to be between 0.03 ng/mL to 0.08 ng/mL.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.