Abstract

The determination of total phosphorus in an automated microbatch analyzer system is described. The procedure combines persulfate digestion of the sample and colorimetric detection of orthophosphate by the ascorbic acid reduction of phosphomolybdate in a single processing chamber. Approximately 9 min are required per sample; ∼ 1.7 ml of sample and 0.17 ml of total reagent are consumed. The limit of detection is ⩽10 μg/1. phosphorus. The method is tolerant of variations in salinity of the sample. Good agreements with ASTM procedures are shown for a variety of test compounds, lake water, wastewater and urine samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.