Abstract

We describe automated column-switching high-performance liquid chromatography for determining nine tricyclic and tetracyclic antidepressants (TCAs) and their metabolites in human serum. TSKgel ODS-80TM and TSKprecolumn PW (Tosoh Co., Tokyo) are used in the analytical column and the precolumn, respectively. A 200-microL serum sample is directly injected onto the precolumn. After washing the serum proteins from the precolumn with potassium phosphate buffer, the precolumn connection is switched to introduce the retained substances onto the analytical column. The drugs are then eluted within 30 min with an acetonitrile/potassium phosphate buffer mixture containing sodium 1-heptanesulfonate. The analytical recoveries (95-104%), reproducibilities (within-run CV less than 3%), and detection limits (10 micrograms/L) indicate that this HPLC system is suited for therapeutic drug monitoring. Correlations were good between the TCA concentrations in serum and administered dose (r = 0.713, n = 41), and between 10-hydroxynortriptyline and nortriptyline in serum (r = 0.691, n = 24).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.