Abstract

The factor V Leiden mutation, a G-->A transition at position 1691 in exon 10 of the gene that codes for factor V, produces an Arg506Gln substitution and is the most common genetic risk factor for venous thrombosis. We have developed a rapid, sensitive, and specific method to detect the factor V Leiden mutation in genomic DNA from whole blood by PCR amplification and microparticle enzyme immunoassay detection using the Abbott LCx instrument. We compared this automated method with the standard procedure using restriction endonuclease digestion of PCR products followed by gel electrophoresis in blinded experiments. In 130 patients (from Veterans Affairs medical centers) with deep venous thromboses, including 24 heterozygotes with the factor V Leiden mutation, there was complete agreement between the two methods. The assay was also able to distinguish heterozygotes from homozygotes. This method, which carries a low potential for cross-contamination of samples, should be a useful routine test for the factor V Leiden mutation in clinical laboratories with sufficient demand for molecular diagnostic assays using the LCx instrument.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.