Abstract

Recently, the XE-2100 hematology analyzer was investigated in a rather small patient group; pseudoeosinophilia or abnormal white blood cell (WBC) scattergrams reported by this instrument were considered as significantly valuable diagnostic parameters in detecting vivax malaria. This study was conducted not only to assess the usefulness of pseudoeosinophilia or abnormal WBC scattergram in vivax malaria-endemic areas with large patient groups (N = 1,801) but also to investigate the correlation of parasitemia and platelet count with pseudoeosinophilia and abnormal WBC scattergrams. Of the 1,801 analyzed patients, 413 (22.9%) were found to have malaria by Wright-Giemsa stained blood smears. Overall, either pseudoeosinophilia or abnormal WBC scattergram was detected in 191 of 413 malaria patients and 4 of 1,388 patients without malaria (sensitivity = 46.2%, specificity = 99.7%). We suggest that clinical hematology laboratories using the XE-2100 analyzer should be aware of such specific parameters, even with the absence of a clinical request.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.