Abstract

Foveal fixation was monitored in normal subjects remotely and continuously by use of a noninvasive retinal scan. Polarized infrared light was imaged onto the retina and scanned in a 3 degrees annulus at 44 Hz. Reflections were analyzed by differential polarization detection. In all 32 eyes studied, the detected signal was predominantly 88 Hz during central fixation (within +/-1 degree) and 44 Hz during paracentral fixation. Phase shift at 44 Hz correlated with the direction of eye displacement. Potential applications of this technique include screening for eye disease, eye position monitoring during clinical procedures, and use of eye fixation to operate devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.