Abstract

This paper presents an automatic method for detection of apnea and hypopnea events occurred during sleep from the single channel recording of oronasal airflow signal. For the identification of events, three time domain measures were extracted from each of the overlapping short segment windows of respiration signal. The feature set includes area, upper 90th percentile and variance, which were used to characterize changes in the airflow signal during normal and abnormal breathing events (i.e., apnea, hypopnea). An ensemble of three binary Support Vector Machine (SVM) based classifiers arranged in one-against-all strategy, were used to classify the feature vector among three categories, according to its origin from some breathing events like normal, apnea and hypopnea. The consecutive decisions of classifier model on time sequenced consecutive overlapped windows were combined by some heuristic rules to identify abnormal breathing events from normal breathings. In this study, 14 polysomnography (PSG) recordings diagnosed as obstructive sleep apnea syndrome were analyzed. Independent test was performed on 6 recordings. The cross-validation and independent test accuracies of apneic event detection were found to be 93.3% and 92.8%, respectively. For hypopnea event these two accuracies were 90.1% and 89.6%. The proposed system can be used for home based monitoring of suspected apneic subject, and can count total number of apnea and hypopnea events occurred during sleep.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call