Abstract

Transmission spectroscopy over large spectral ranges (>100 cm(-1)) generally requires a reference measurement to be taken separately from the sample scan. The ratio of the two measurements (i.e., the transmittance) is therefore susceptible to baseline changes that occur between the recording of the two spectra. The origins of relatively strong baseline changes (≫1%) of a difference-frequency-generation-based laser spectrometer (tuning range 2900-3144 cm(-1), 150 μW average power) were investigated and a method for minimizing them by improving reproducibility and reducing measurement time is presented. The new method was tested for a gas mixture and the sensitivity for broad absorption features was determined as 5×10(-3) minimum measurable absorbance for a total scan duration of 70 min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.