Abstract

We describe a fully automated method of analysis for plasma epinephrine and norepinephrine by use of a two-column system of "high-performance" liquid chromatography. Catecholamines in deproteinized plasma are purified on the first (preparation) column, then transferred automatically to the second (analytical) column in which epinephrine and norepinephrine are resolved. These compounds are then determined fluorometrically with a continuous-flow reaction system by the trihydroxyindole method. The minimum sensitivity of the method is 0.02 and 0.04 pmol for epinephrine and norepinephrine, respectively. One assay can be completed in 30 min, and greater than 1 mL of plasma is required for the procedure. The within-run CV in the chromatographic determination of pooled plasma was greater than 3%, and the analytical overall recovery of the two compounds was 95%. The concentrations in plasma of medical students at rest were 0.32 (SE 0.06) nmol/L for epinephrine and 0.98 (SE 0.08) nmol/L for norepinephrine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.