Abstract

An unconventional autocorrelation method is described for measuring the transfer function of optical systems. The interference takes place between the scattered waves obtained from two laterally sheared correlated partial diffusers. The output of a detector responding only to an extremely-narrowband of spatial frequencies is proportional to the autocorrelation of the system pupil function. An automatic display of the transfer function is obtained by continuously varying the shear between the diffusers. We present the theory and some experimental results of this simple and inexpensive device. A study of various parameters affecting the performance of the instrument is also given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.