Abstract
The silicate (Si) molybdenum blue method was modified by combining oxalate and ascorbic acid into a single reagent and was used for determining Si in sea water samples. The first step of this automated assay protocol was designed to perform either a calibration by a single Si standard prepared in deionized (DI) water, or to dilute samples in the range of 0–160 μM Si to fit into 0–20 μM Si calibration range using a 20 cm flow cell. By designing the assay protocol to function in batch mode, the influence of salinity on calibration was eliminated, thus making the method suitable for analysis of samples collected in the open ocean, coastal areas, or rivers. Reproducibility and accuracy of this method were evaluated by analysis of certified sea water reference materials. Phosphate (P) does not interfere significantly if the Si:P ratio is 4:1 or larger. The limit of detection was 514 nM Si, r.s.d. 2.1 %, sampling frequency 40 s/h, reagent consumption 700 μL/sample, and using deionized water as the carrier solution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.