Abstract

The performances of YAG(Ce) and CsI(Tl) scintillators have been compared using photomultiplier tube (PMT) readout for γ rays and α particles. It is found that the energy resolution of YAG(Ce) is inferior to that of CsI(Tl). With Philips XP2971 PMT, we have obtained improved energy and time resolution for YAG(Ce), compared to the same, obtained by earlier workers using XP2020/Q. The best values of energy resolution (FWHM), obtained in the present work, for 662keV γ ray, are 6% and 7%, respectively, for CsI(Tl) and YAG(Ce), whereas for 5.48MeV α particles, the observed values are 6% and 8.4%, respectively. The pulse height response of both the scintillators has been found to be reasonably linear up to 1.3MeV γ energy. A prompt time resolution of 1.3ns (FWHM) has been obtained with a BaF2 – YAG(Ce) combination against 511–511keV photopeak selection, which compares well with that obtained for a BaF2–CsI(Tl) combination, used by us.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.