Abstract
The aim of this study was to investigate the ability of Australian centres to produce high-dose intensity modulated radiation therapy (IMRT) prostate plans, and to compare the planning parameters and resultant dose distributions. Five Australian radiation therapy departments were invited to participate. Each centre received an identical 5 mm-slice CT data set complete with contours of the prostate, seminal vesicles, rectum, bladder, femoral heads and body outline. The planning team was asked to produce the best plan possible, using published Memorial Sloan-Kettering Cancer Centre prescription and dose constraints. Three centres submitted plans for evaluation. All plans covered the planning target volume adequately; however, only one plan met all the critical organ dose constraints. Although the planning parameters, beam arrangements and planning systems were different for each centre, the resulting plans were similar. In Australia, IMRT for prostate cancer is in the early stages of implementation, with routine use limited to a few centres.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.