Abstract
Background. Auscultatory measurement using a sphygmomanometer has been the predominant method for clinical estimation of blood pressure, but it is now rapidly being replaced by oscillometric measurement. Objective. To compare blood pressure by auscultatory and oscillometric measurements in patients ≥ 80 years. Method. 100 patients had blood pressure measured by auscultation with a sphygmomanometer and by an electronic device using the oscillometric method. For each patient the mean of two blood pressures with each method measured within 15 min were compared. Results. The mean age of participants was 85.8 years; 55.8% were women. The correlation coefficient for systolic blood pressure was 0.88 and for diastolic 0.79. Differences between auscultatory and oscillometric values were less than 10 mmHg in 70.6% of systolic blood pressures and in 83.2% for diastolic. Arrhythmia and hypertension did not influence the results, and there was no correlation between the magnitude of the differences and the level of blood pressure. Conclusion. Agreement between oscillometric and auscultatory measurements of blood pressure in octogenarians was found to be less than required by validation protocols. However, semi-automatic equipment, which is observer-independent, may be used even in the very elderly, particularly if multiple readings are performed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.