Abstract

Orthopaedic screws insertion can be trivialised as a simple procedure, however it is frequently performed poorly. Limited work exists defining how well surgeons insert screws or whether augmented screwdrivers can aid surgeons to reduce stripping rates and optimise tightness. We aimed to establish the performance of surgeons inserting screws and whether this be improved with screwdriver augmentation. 302 orthopaedic surgeons tightened 10 non-locking screws to what they determined to be optimum tightness into artificial bone sheets. The confidence in the screw purchase was given (1–10). A further 10 screws were tightened, using an augmented screwdriver that indicated when a predetermined optimum tightness was reached. The tightness for unstripped insertions under normal conditions and with the augmented screwdriver were 81% (95% CI 79–82%)(n = 1275) and 70% (95% CI 69–72%)(n = 2577) (p < 0.001). The stripping rates were 58% (95% CI 54–61%) and 15% (95% CI 12–17%) respectively (p < 0.001). The confidences when using the normal and augmented screwdrivers respectively were 7.2 and 7.1 in unstripped insertions and 6.2 and 6.5 in stripped insertions. Performance improved with an augmented screwdriver, both in reduced stripping rates and greater accuracy in detecting stripping. Augmenting screwdrivers to indicate optimum tightness offer potentially enormous clinical benefits by improving screw fixation.