Abstract

IntroductionThe periacetabular osteotomy is a technically demanding procedure with the goal to improve the osseous containment of the femoral head. The options for controlled execution of the osteotomies and verification of the acetabular reorientation are limited. With the assistance of augmented reality, new possibilities are emerging to guide this intervention. However, the scientific knowledge regarding AR navigation for PAO is sparse.MethodsIn this cadaveric study, we wanted to find out, if the execution of this complex procedure is feasible with AR guidance, quantify the accuracy of the execution of the three-dimensional plan, and find out what has to be done to proceed to real surgery. Therefore, an AR guidance for the PAO was developed and applied on 14 human hip cadavers. The guidance included performance of the four osteotomies and reorientation of the acetabular fragment. The osteotomy starting points, the orientation of the osteotomy planes, as well as the reorientation of the acetabular fragment were compared to the 3D planning.ResultsThe mean 3D distance between planned and performed starting points was between 9 and 17 mm. The mean angle between planned and performed osteotomies was between 6° and 7°. The mean reorientation error between the planned and performed rotation of the acetabular fragment was between 2° and 11°.ConclusionThe planned correction can be achieved with promising accuracy and without serious errors. Further steps for a translation from the cadaver to the patient have been identified and must be addressed in future work.