Abstract
PurposeTo illustrate a surgical technique for augmented reality (AR)-assisted unicompartmental knee arthroplasty (UKA) and report preliminary data.MethodsWe developed an AR-based navigation system that enables the surgeon to see the tibial mechanical axis superimposed on the patient’s leg in addition to the tibial cutting angle. We measured the tibial resection angle in 11 UKAs using postoperative radiographs and calculated the absolute difference between preoperative target angle and postoperative measured angle. The target angle was determined for each patient: mean values were 0.7° ± 1.0° varus in coronal alignment and 5.3° ± 1.4° posterior slope in sagittal alignment.ResultsThe angles measured on postoperative radiographs were 2.6° ± 1.2° varus in the coronal plane and 4.8° ± 2.5° posterior slope in the sagittal plane. The absolute differences between the target and measured angles were 1.9° ± 1.5° in coronal alignment and 2.6° ± 1.2° in sagittal alignment. No patients experienced complications, including surgical site infection and periprosthetic fracture.ConclusionThe AR-based portable navigation system may provide passable accuracy in terms of proximal tibial resection during UKA.Level of EvidenceIV
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.