Abstract

A degree of judgement is required when applying the British Standard associated with the calibration and performance of audiometers, BS EN 60645-1 (IEC 645-1). Consequently there is potential for variation of practice between calibration agencies. This note addresses issues and makes suggestions relating to the measurement of distortion, on-off effect, cross talk, linearity, rise/fall time and to the positioning of the earphone on an artificial ear. The degree of precision appropriate to each measurement is suggested. A warning is given of the audiological consequences of excessive distortion produced by normally functioning audiometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.