Abstract

Significantly improved sensitivity for analysis of biomolecules by MALDI FT-ICR mass spectrometry is achieved by (i) microscope-monitored sample deposition onto a small indentation on the probe tip and (ii) multiple remeasurement of ions from a single laser shot. A simple modification to the solids probe tip allows for microdeposition of a few amols of analyte onto small indentation spots previously aligned with the laser beam. Ion multiple remeasurement of the same ion packet enhances the signal-to-noise ratio and thus extends the achievable FT-ICR MS detection limit. We demonstrate that FT-ICR can be used to detect parent and structurally significant fragment ions of peptides and phospholipids at low amol amounts. Positive ion mass spectra for approximately 90 amol of a mixture of angiotensin II and bradykinin, approximately 40 amol of dipalmitoylglycerophosphatidylcholine, and approximately 8 amol of substance P constitute the lowest reported detection limits to date for FT-ICR mass analysis of MALDI-generated ions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.