Abstract

Single- and three-phase broad-beam x-ray attenuation data have been obtained using lead, steel, plate glass, gypsum wallboard, lead acrylic, and wood. Tube voltages of 50, 70, 100, 125, and 150 kVp were employed and the resulting curves were compared to transmission data found in the literature. To simplify computation of barrier requirements, all data sets were parametrized by nonlinear least-squares fit to a previously described mathematical model. High attenuation half value layers and the lead equivalence of the alternate materials were also determined.