Abstract
The fracture of the femoral component is a rare complication of a total knee arthroplasty (TKA). This article presents a case in which a 70-year-old man underwent a left press-fit Advance Medial-Pivot Knee System TKA (MicroPort Orthopedics, Arlington, TN) in 2000. Twenty years later, he experienced a nontraumatic onset of knee pain after standing up from a lunge position. Radiographs and CT scans revealed a complete fracture of the medial condyle of the femoral component. Revision surgery was performed confirming the broken component. A cemented Triathlon Total Stabilizer (Stryker Orthopedics, Kalamazoo, MI) prosthesis was used for the revision. The authors recommend that surgeons maintain a high level of suspicion of component fracture when patients present with persistent severe knee pain and instability after a TKA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.