The Canadian Penning Trap mass spectrometer was designed to determine precisely the masses of stable and unstable isotopes. To date, such measurements have been carried out on approximately 60 short-lived species. A laser ablation ion source is also available to produce ions of stable isotopes, intended for use in calibrations, checks for systematic effects and for measurements involving stable isotopes. Mass ratios for the isotopes 194,195,196,198Pt relative to 197Au have been determined to a precision of better than 3 × 10-8. These measurements were motivated, in part, by the long-standing discrepancy between earlier mass measurements and the Atomic Mass Evaluations in the mercury region. The results also demonstrate the stability of the measurement system and set limits on the magnitude of systematic effects. No significant deviations from accepted values were found.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.