Abstract

We constructed an optical pickup head atomic force microscope (OPH-AFM). The component used to monitor the displacement of the cantilever is made from the optical pickup head of a commercial DVD-ROM. In contrast to previous trials using an optical pickup head, this OPH-AFM provides a direct view of the cantilever and the sample. Therefore, the preliminary imaging process, including placing the laser beam on the cantilever and setting the initial scanning position over the sample, is straightforward. The OPH-AFM showed vertical resolution of below 20 nm, which is close to the maximum performance of the related hardware.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call